The clinical investigation is conducted as a controlled (subject as own control), prospective, randomised, comparative investigation in a population with a representative spread of BPs to ensure that the ID performs as good as current marketed and approved devices. The study will be conducted in compliance with the International Standard ANSI/AAMI/ISO 81060-2 - Non-invasive sphygmomanometers - Clinical validation of automated measurement type.
This clinical investigation will enrol up to 96 male and female subjects to ensure the recording of 85/35 evaluable datasets in accordance with the ANSI/AAMI/ISO standard \[11\]. All measurements will be paired and falls into 2 categories: Auscultatory measurements as gold standard reference, oscillometric measurements for clinical practise reference and further validation of ambulatory monitoring and pulse shape measurements.